The Foxy R2 fraction collector can be adapted to a broad spectrum of applications.
The Foxy R2 doubles the collection capacity, and uses Radio Frequency Identification (RFID) for error-free rack configuration.
Key features
- Fraction collector for two racks
- Collect uniform fractions by time or pumped volume.
- Collect peaks of interest separately using any combination of slope, level, and time.
- Stores up to 8 collection methods.
- Choice of available racks
- Coldroom-storage possible.
- Easy to use symbolic touch screen interface
- Fast tube changes
- Small delay volume between diverter valve and drop former
- Flow rates up to 125 ml/min
| Fraction collection | |
|---|---|
| Number of racks |
2 |
| RFID rack recognition |
yes |
| Capillary connection |
1/8″ |
| Fractionation modes |
drop counting, time intervals, volume intervals |
| Fractionation flow rate |
standard max. 125 ml/min
|
| Fraction capacity |
4 x 96-well microplates 1.5 ml Eppendorf tubes (60/rack) 2 x 12 mm test tubes (144/rack) 2 x 13 mm test tubes (144/rack) 2 x 16 mm test tubes (100/rack) 2 x 18 mm test tubes and 15 ml Falcons (72/rack) 2 x 25 mm test tubes (36/rack) 2 x 50 ml Falcons (36/rack) 2 x 480 ml bottles (9/rack) Funnel racks for fractions larger than 480 ml |
| Diverter valve |
drop former (NC): 110 µl waste (NO): 130 µl |
| Wetted materials |
valve: PEEK and perfluoroelastomer (FFKM), supplied ferrules: ETFE, supplied valve tubing: PTFE, supplied drain tubing: viny |
| Fractionation control |
operator: front panel control via touch screen LCD integrated systems: direct communication via Ethernet (TCP/IP) and RS-232 serial communications |
| Maximum test tube height |
190 mm |
| Communication | |
| Control |
LAN, RS-232 |
| Technical parameters | |
| Conformity |
CE, CSA |
| Display |
touch screen LCD displays |
| Ambient conditions |
0–40 °C, 32–104 °F |