AZURA P 6.1L Ultra High Pressure Pump with 5 ml pump head, stainless steel
Technical Data
| Max. flow rate |
5 ml/min |
| Flow rate range |
0.001 – 5 ml/min |
| Flow rate increment |
0.001 ml/min |
| Maximum delivery pressure [psi] |
14500 psi |
| Maximum delivery pressure [bar] |
1000 bar |
| Maximum delivery pressure [MPa] |
100 MPa |
| Pressure range |
|
| Pump head materials |
Stainless Steel |
| Gradient |
High Pressure Gradient (binary) |
| Leak management |
Yes |
| Wetted materials |
GFP (graphite fiber reinforced PTFE), sapphire, ruby, aluminum oxide (Al2O3) |
| Best working conditions |
0.02 – 5 ml/min |
| Continous working conditions |
0.1 – 4.0 ml/min |
| Flow rate accuracy |
< 1 % |
| Flow rate accuracy conditions |
measured at 5-80 % of flow range, using ethanol |
| Flow rate precision |
0.1 % RSD |
| Flow rate precision conditions |
based on the retention time at constant room temperature |
| Gradient range |
0 – 100% |
| Gradient range (recommended) |
5 – 95 % |
| Gradient accuracy |
± 0.3 % at 1 ml/min, 150 bar (ethanol/caffeine tracer) ± 1 % (5 – 95 %, measured at 0.1 – 10 ml/min, water/caffeine tracer) |
| Gradient precision |
< 0.1 % RSD at 1 ml/min, 0.3% RSD overall, based on retention time at constant room temperature |
| Gradient minimum increment |
0.1 % |
| Mixing volume |
100 µl |
| Delay volume |
110 µl with 100 µl Mixer |
| Pump type |
Binary analytical HPLC pump with degasser |
| Degasser |
Yes |
| Solvent selection valve |
2 x 2 channels |
| Pulsation |
< 2 % Amplitude (typically: < 1,3 %) or < 3 bar (0,3 MPa), whatever is greater, at 1 ml/min ethanol, at all pressures > 10 bar (1 MPa, 147 psi) |
| Pulsation compensation |
Active pressure and pulsation compensation |
| Active piston seal wash |
Yes |
| Piston seal washing |
Yes |
| System protection |
Soft start, Pmin und Pmax are programmable |
| Pump head inlet (standard) |
UNF 1/4-28 Thread |
| Pump head outlet (standard) |
UNF 10-32 Thread |
| Interfaces |
LAN, Pin header connectors (Analog IN, Start In, Error IN) |
| Control |
LAN, Analog and event control, Mobile Control |
| Analog inputs |
Flow rate, 0 – 10 V via pin header connectors |
| Analog outputs |
8 event outputs (TTL, OC, Relais) and 24 V |
| Power supply |
100 – 240 V; 50 – 60 Hz; Maximum power consumption 100 Watt |
| Dimensions |
361 mm x 208.2 mm x 523 mm (W × H × D) |
| Weight |
14.1 kg |
| Leak sensor |
Yes |
| Ambient conditions |
4-40 °C (39.2-104 °F) Air humidity below 90%, non-condensing |
-
AZURA Pump P 6.1L User Manual EN (V6890)
-