ANSER MWD-2003F
Technical specifications
| Instrument type | Double-beam photometer |
| Light source | Xenon flash lamp, Fiber Optics |
| Wavelength range | 190 – 750 nm |
| Number of wavelength channels | 1 – 4 variable |
| Wavelength accuracy | ± 1.0 nm
|
| Wavelength precision (reproducibility) | ± 0.5 nm |
| Spectral resolution | <6 nm at 254 nm |
| Noise
|
<± 1.0 x 10-3 AU (dry analytical cell, 255 nm, 1s rise time) |
| Drift | <± 1 x 10-3 AU/h (dry analytical cell, 255 nm, 1 hour after power-up) |
| Linearity range | > 2.5 AU |
| Response time | 0.0 / 0.02 / 0.05 / 0.1 / 0.2 / 0.5 / 1.0 / 2.0 s |
| Output data rate | 1 / 2 / 5 / 10 / 20 / 50 / 100 Hz |
| Communications | – LAN 10 / 100 Mbit/s
– USB (virtual COM port) |
| Stand-alone operation | TFT touch panel |
| Remote operation | Autozero digital input
Start digital input Analog input to control wavelength Analog output for signal |
| Flow cell | Remote, fiber optic connectors |
| Power | External power supply, 24 Vdc, 1 A max. |
| Dimensions | 95 x 160 x 280 mm / 3.75 x 6.5 x 11.0 in.
(W x H x D) |
| Weight | 2 kg / 4,5 lbs. |