ANSER VWD-1001F

Product No:

VWD-1001F

ANSER VWD- 1001F
Photometric Detectors

UV-Detector with Deuterium lamp light source, variable wavelength (UV range 190-420 nm)

Technical specifications

Instrument type Double-beam spectrophotometer
Light source Deuterium lamps
Wavelength range UV/VIS, 190 – 420 nm
Number of wavelength channels 1 variable
Wavelength accuracy ± 1.0 nm

 

Wavelength precision (reproducibility) ± 0.5 nm
Spectral resolution <6 nm
Noise

 

<± 0.3 x 10-5 AU (dry analytical cell, 240 nm, 1s rise time, single wavelength)
Drift <± 1 x 10-4 AU (dry analytical cell, 240 nm, 1s rise time,
single wavelength)
Linearity range > 2.5 AU
Response time 0.0 / 0.02 / 0.05 / 0.1 / 0.2 / 0.5 / 1.0 / 2.0 s
Output data rate 100 Hz (1 wavelength)
Communications LAN 10 / 100 Mbit/s
USB 2.0
RS-232-C
Stand-alone operation 4“ LCD touch panel
Remote operation Start/Stop input
Autozero input
Error output
Wavelength tuning analogue input, 0-10V
2x Absorption signal analogue outputs, ±2.5 V, 24 bits resolution
Flow cells 10 mm path length, 2.5 μl volume (micro)
10 mm path length, 8 μl volume (analytical)
0.5 – 2 mm path length, 2 – 8 μl volume (preparative)
Other flow cells upon request
Cell material: stainless steel, titanium, PEEK
Fiber optic flow cell
Power External power supply, 24 Vdc, 2.5 A max.
Dimensions 127 x 178x 350 mm / 5.0 x 7.0 x 13.7 in.

(W x H x D)

Weight 5 kg / 11 lbs.
Data communication LAN 10 / 100 Mbit/s USB 2.0 RS-232-C (115200 baud)