Variable single wavelength with fiber optic connectors
The AZURA UV/VIS Detector UVD 2.1L is a competitively priced HPLC spectrophotometer for routine HPLC applications including fast LC methods. Besides offering excellent technical specifications, this robust detector features a highly flexible and compact design. The UVD 2.1L comes with an installed deuterium lamp which covers a wavelength range from 190 to 750 nm. This detector can be controlled with OpenLab® EZChrom Edition, ChromGate® and ClarityChrom® software, as well as from an optional touch display (stand-alone operation), via LAN or through analog input/output; allowing it to be integrated into almost any LC system. Due to a smart design the flow cell is easily accessible and can be changed very quickly. You can choose between a wide range of flow cells for analytical or preparative LC applications with flow rates from 10 μl/min up to 10 l/min. This fiber optics version offers the possibility to separate the flow cell spatially from the device and thus provides enhanced security for hazardous, explosive or toxic work processes.
Technical Data
| Detector type |
Variable single wavelength UV detector |
| Version |
fiber optics version |
| Detection channels |
1 |
| Light source |
Deuterium (D²) lamp with integrated GLP chip |
| Wavelength range |
190 – 750 nm |
| Spectral bandwidth |
11 nm at Hα line (FWHM) |
| Wavelength accuracy |
± 2.5 nm |
| Wavelength precision |
0.3 nm (ASTM E275-93) |
| Noise |
± 2.0 x 10-5 AU at 254 nm (ASTM E1657-98) |
| Drift |
4.0 x 10-4 AU/h at 254 nm (ASTM E1657-98) |
| Linearity |
> 2.0 AU at 270 nm (ASTM E1657-98) |
| Maximum data rate |
50 Hz (LAN)/20 Hz (Analog) |
| Flow cell |
Not included (see Accessories / Spare parts) |
| Time constants |
0.0 / 0.1 / 0.2 / 0.5 / 1.0 / 2.0 / 5.0 / 10.0 s |
| Integration time |
Automatic |
| Leak sensor |
Yes |
| Inputs |
Error (IN), Start (IN), Autozero,0 – 10 V Analog IN |
| Outputs |
Events 1 – 3, + 5 V, 24 V Valve |
| Analog outputs |
1 x 0 – 5 V scalable, 20 bit, offset adjustable |
| Control |
Digital: LAN, remote connector/Analog: wavelength control/Manual: Mobile Control (optional) |
| Programming |
Timed: wavelength, events, fraction valve, links, wake up (program, link); /9 programs, 50 program lines |
| GLP |
Detailed report including lamp recognition, operating hours, /lamp operating hours, number of lamp ignitions |
| Display |
Mobile Control (optional) |
| Ambient conditions |
Temperature range: 4 – 40 °C, 39.2 – 104 °F, Humidity: below 90 % noncondensing |
| Power supply |
100 – 240 V, 50 – 60 Hz, 75 W |
| Dimensions |
361 x 158 x 523 mm (W x H x D) |
| Weight |
6.0 kg |
| Note |
Fiber optic cables and flow cells are not included and must be ordered/ separately (see Accessories / Spare parts) |
-
AZURA Detektor UVD 2.1L Benutzerhandbuch DE (V6831)
-
AZURA Detector UVD 2.1L User Manual EN (V6831)
-
List of supported devices in chromatography data systems EN (V1663)
Device support in KNAUER CDS / KNAUER device support in 3rd party software