| Delivery System |
Double-piston pump |
| Maximum Operating Pressure |
100 mL head 400 bar
250 mL head 200
bar
500 mL head 100 bar
1,000 mL head 50 bar |
| Flow Rate Range |
100 mL head
0.1 - 99.9 mL / min
250
mL head 0.1 - 249.9 mL / min
500 mL head 0.1 - 499.9 mL / min
1,000 mL head 0.1 - 999.9 mL / min |
| Flow Rate Precision |
± 2% within 2 - 50 % of flow range,
using Methanol / H2O 80/20 v/v |
| Reproducibility |
< 0.1 % |
| Pulsation |
Electronic compensation of residual
pulsation |
| System Protection |
Soft start, Pmin and Pmax adjustable |
| Control |
2 RS232 ports
8 event outputs (TTL, OC, relays)
Remote connections (start, error)
Analog pressure monitor, 2 analog inputs
(flow rate), Expandable to high-pressure gradient system with up to 4 eluents;
Digital
remote control is supported by Knauer's software package EuroChrom 2000 preparative
version, enabling full and transparent integration of the K - 1800 pump into a wide range
of chromatographic systems. |
| Power Supply |
115 / 230 V, 47 - 63 Hz, 450 W |
| Gradient |
High pressure gradient: configurable
master pump system with up to 3 slave pumps
Low pressure gradient upon request |
| Dimensions and weight |
220 x 185 x 340 mm ( W x H x D ) 10.1
kg |